FA 515 EX Intrinsically Safe Stationary Dew Point Transmitter
The newly developed FA 515 Ex intrinsically safe dew point transmitter by CS Instruments (Germany) performs pressure dew point measurements in potentially explosive atmospheres and is suitable for non-corrosive gases.
Typical Applications: Air / compressed air, argon (Ar), nitrogen (N2), biogas, natural gas, hydrogen (H2), etc.
The explosion-proof FA 515 Ex dew point sensor complies with the following industry certifications:
ATEX II 2G Ex ib IIC T4 Gb (Zone 1) Gas, intrinsically safe, max. surface temperature 135 °C
ATEX II 2D Ex ib IIIC T80°C Db (Zone 21) Dust, intrinsically safe, max. surface temperature 80 °C
Product Application
-
Compressed Air Systems
Application: Monitoring dew point temperature in dryers (desiccant, refrigerated). -
Drying Process Monitoring
Industries: Dew point monitoring for plastic injection molding, food processing, pharmaceuticals, and electronics manufacturing. -
Gas Purity Dew Point Monitoring (e.g., N2, H2, O2)
Industries: Semiconductor, chemical, laser, and medical applications. -
Cold Storage & Cold Chain Logistics
Electrical Power Equipment (SF6 Switchgear Gas Monitoring)
Application: Moisture in SF6 gas generates acidic byproducts that degrade electrical insulation equipment. -
Climatic & Environmental Simulation Testing
Applications: Laboratories, automotive component testing, and military equipment environmental durability testing.
FA 515 EX Dew Point Transmitter Key Advantages
-
Suitable for all types of dryers with a measuring range from -100 to +50 °Ctd dew point.
-
Gas-independent measurement, also suitable for other non-corrosive gases.
-
Compact design, ideal for installation in dryers.
-
Robust design with IP66 protection class.
-
Long-term stability and drift resistance, optimizing calibration intervals.
-
Analog output 4...20 mA (2-wire or 3-wire technology) and Modbus RTU (RS 485) interface.
-
Available in multiple measurement units: pressure dew point [°Ctd], temperature [°C], relative humidity [%RH], absolute humidity [g/m³], moisture content [g/kg], moisture fraction V/V [ppmV/V], water vapor partial pressure [hPa], and atmospheric dew point [°Ctd.atm].
FA515 EX Dew Point Transmitter Technical Data
-
Measuring Range: -100 to +50 °Ctd
Dew Point Accuracy:
±1 °Ctd (-20 to +20 °Ctd)
±2 °Ctd (-50 to -20 °Ctd)
±3 °Ctd (-80 to -50 °Ctd)
Operating Pressure: -1 to 50 bar
Operating Temperature: -20 to +70 °C
Power Supply: 24 VDC
Analog Output: 4–20 mA (2-wire) = -20 to +50 °Ctd
Digital Interface: Modbus RTU (RS-485)
Process Connection: G 1/2" thread
Dimensions: Ø 30 mm, L 130 mmATEX II 2G Ex ib IIC T4 Gb (Zone 1) Gas, intrinsically safe, max. surface temperature 135 °C
ATEX II 2D Ex ib IIIC T80°C Db (Zone 21) Dust, intrinsically safe, max. surface temperature 80 °C